Cobra-S 800 Series

High resolution, high speed & proven imaging performance

Launched in 2017, the Cobra-S is our original and most popular high-speed spectrometer for near infrared OCT laboratory research. It is used in research applications in ophthalmology and materials processing as well as more specialized techniques like polarization-sensitive OCT (PS-OCT) and optical coherence elastography (OCE). It is cited widely in literature, and we continuously marvel how this enabling technology is helping academics push the boundaries of OCT research.

  • Imaging option from 400 – 700 nm; collect blood perfusion data and ultra-high resolution OCT images simultaneously at 550 nm
  • Low crosstalk camera & diffraction-limited optics for exceptional roll-off; as little as 6 dB at 1.2 mm 
  • Proprietary VPH gratings; highly efficient optical design with superior SNR & subpixel resolution 
  • USB 3.0 interface option for 20-130 kHz models allows easy, cost-effective system integration 
  • OEM ready: robust, compact & athermal
CS800-930/60
CS800-840/120
CS800-850/140
CS800-840/180
CS800-700/160
CS800-780/200
CS800-800/300
Wavelength Range
900–960 nm
780–900 nm
780–920 nm
750–930 nm
620–780 nm
680–880 nm
650–950 nm
Bandwidth
60 nm
120 nm
140 nm
180 nm
160 nm
200 nm
300 nm
Imaging Depth
7.4 mm
3.0 mm
2.5 mm
2.0 mm
1.5 mm
1.5 mm
1.0 mm
Spectral Resolution
0.03 nm
0.06 nm
0.07 nm
0.09 nm
0.08 nm
0.10 nm
0.15 nm
Axial Resolution
6.3 µm
2.6 µm
2.3 µm
1.7 µm
1.3 µm
1.3 µm
0.9 µm
Max Line Rate
20 kHz, 80 kHz, 130 kHz, or 250 kHz
Pixels
2048
Interface
USB 3.0 (up to 130 kHz) or Camera Link (up to 250 kHz)
Dimensions
9.5 cm x 19 cm x 6 cm
Weight
1.4 kg
Cobra-S 800 part numbers are of the format CS800-[CWL]/[BW]-[kHz]-[camera][pixels]-[interface], where the e2v OctoPlus camera is represented by “OC”, 2K denotes 2048 pixels, and the interface is either camera link (up to 250 kHz, denoted by “-CL”) or USB 3.0 (up to 130 kHz, , denoted by “-U3”). Imaging depth stated is the theoretical value, calculated in air.
  • USB-3.0: 20kHz, 80kHz, 130kHz
  • Camera Link: 250kHz

Like our flagship Cobra-S 800, the Cobra VIS OCT spectrometer offers superior roll-off performance, as can be seen in this comparison of signal with imaging depth. With less roll-off, much deeper structures can be imaged clearly than is expected of typical 800 nm imaging.

  • Angiography
  • 3D blood oxygenation mapping
  • Explore pathology & structure
  • High-res retinal imaging
  • 3D microscopy

Compatible with Wasatch Photonics’ SDKs for rapid data access and control in development environments (e.g., C++, LabVIEW, MATLAB).

Compatible with Wasatch Photonics’ SDKs for rapid data access in development environments (e.g., C++, LabVIEW, MATLAB).